Low speed large capacity centrifuge operation and matters needing attention

Release date:2014-11-21

On low speed large capacity centrifuge the other information please visit http://www.shlxj.com.cn/ for more details.
A low speed, large capacity centrifuge operation procedure:
1 plug in the power, standby indicator lights: turn the power switch, the flashing digital tube digital speed regulation and timing system appears to factory of machine working speed setting.
The 2 set of low speed large capacity centrifuge machine working parameters, such as temperature, operation time, working speed etc.. Operation control button on the panel, the centrifuge is running. In accelerating the predetermined time, the rotational speed rises to the preset value. Low speed large capacity centrifuge at a predetermined operation time (not including the deceleration time), the beginning of the centrifuge deceleration, speed down to zero in the deceleration time preset in. According to the control stop button on the panel, digital display dedT, after a few seconds that appears flashing speed value, this opportunity is ready for the next work.
Two, low speed and large capacity centrifuge note:
1 low speed large capacity centrifuge in addition to running speed and running time, working parameters do not change the machine at random, so as not to affect the performance of the machine. Speed setting shall not exceed the maximum speed, to ensure the safe operation of the machine. Use as presented in 0 or other numbers, the machine is not running, should be shut down power, 10 seconds after a reboot, to be the set speed show, then press the run button, low speed and large capacity centrifuge will run as usual. Not in the low speed large capacity centrifuge machine in operation or rotor without stop condition opened the door. The centrifuge must equal perfusion samples, cut do not make head operation in unbalanced situation. Centrifuge once it is best not to exceed 60 minutes. The centrifuge must be reliable grounding; the machine does not use, should unplug the power plug.